Multi-Purpose Flow Cell (MPFC)

Features

Multi-Purpose Flow Cell (MPFC) Options

  • Suitable for many installations in side streams or sample conditioning systems
  • Collimated beam for excellent light transmission
  • Clean out port for easy window cleaning
  • Compatible with all Guided Wave analyzers

Talk to us about your particular process installation needs for the best fit.

Specifications
Specifications Multi-Purpose Flow Cell
Uses / Applications Most process streams including many with acids, bases, or cyanides
Pressure Range (psi) 0 to 500
Maximum Temperature (°C) 300 (o-ring dependent)
Spectral Range (nm) UV-Vis (200-850); Vis-NIR (400-2100)
Optical Pathlength (mm) 1;2;5;10;20
Connections 3/8in tubulation (for pathlengths 1-5mm)
1/2in tubulation (for 10mm pathlength)
1in tubulation (for 20mm pathlength)
Standard Material 316L SS
Window Material Sapphire (Vis-NIR) or Fused Silica (UV)
Window Seal Polymer o-ring materials: Viton, Kal-Rez® 6375, others on request
Fiber Connector SMA 905
Mounting 3/4in MNPT Conduit connection standard (gland connection also available)
Part Number 45200-xxxxx; 45300-xxxxx
Data Sheet Doc # download 1019
Options

Multi-Purpose Flow Cell (MPFC) Options

The Multi-Purpose flow cell is available with several options for easy adaptation to different process installation requirements.

Heated Flow Cell (PN 45301-xxxxxx) ►

  • Flow cell has a threaded chamber for circulating a heated (or cooled) fluid

CRN (Canadian Registry Number) certified versions (rated to 1000psi) are available

NACE MR0103 material certifications are available

Turbidity Flow Cell (PN 45302-xxxxx) ►

  • 90 degree backscatter measurement
  • requires a 3rd fiber optic cable for the 90 degree measurement
  • for use with the ClearView® db configured for Haze or Turbidity

Teflon Flowcell (PN 12680-xxxxx) ►

  • For applications requiring no metal contact to the sample stream

Teflon PEEK Flowcell (PN 45053-xxxxx) ►

  • For applications requiring no metal contact to the sample stream or exposed metal parts
  • Good for corrosive environments

High Safety Flowcell (PN12526-xxx) ►

  • Dual sealed for safety critical applications
  • Can be set up for leak detection between the primary and secondary sealing surfaces

Axial Flowcell (PN45004-xxxxx) ►

  • For pathlengths over 20mm
Contact Us

We're not around right now. But you can send us an email and we'll get back to you, asap.