The NIR-O Full Spectrum Analyzer is a Dispersive NIR spectrometer with an IECEX/ATEX certified designed for Class 1 Div 1 and Div 2 environments. Every NIR-O comes equipped with Modbus for communicating with a distributed control system (DCS). For customers that require 4-20mA outputs, or other analog or digital, I/O Guided Wave has engineered an optional accessory I/O enclosure.
The M508plus UV-VIS Process Spectrometer is an EX IIC T4 compliant design developed for the polymer industry. The IP65 rated enclosure can include up to 16 custom digital and analog I/O and Modbus. There is no need for an external accessory enclosure.
How Many Channels Are Included on Each Module?
Each Digital Input module can accept up to 4 input channels.
- If you require only 1 digital input, then you need 1 Digital Input module. You will then will have the ability to add up to 3 more inputs in the future.
Each Digital Output module can process to 4 answers or outputs.
- If you only have 2 measurement parameters, then 1 Digital Output module will meet your needs and allow up to 2 more outputs in the future.
Each Analog Input module can accept up to 2 input channels.
- If you require only 1 analog input, a Wika RTD to monitor process temperature, then 1 Analog Input module will provide the ability to add up to 1 more input in the future.
Each Analog Output module can process up to 2 answers or outputs.
- If you have 3 measurement parameters, then you need 2 Analog Output modules.
Configuring the I/O Part Number
The Fiber Glass Cabinet – I/O Enclosure includes the programmable logic controller, 16 position rack for I/O modules, and two power supplies (5V and 24V). The smart part number 40217-1XADO
X = the number of Analog Input modules
A = the number of Analog Output modules
D = the number of Digital Input modules
O = the number of Digital Output modules
PART NUMBER | MODULES INCLUDED |
40217-10133 | 1 x Analog Output, 3x Digital Inputs, 3x Digital Outputs |
40217-10101 | 1 x Analog Output, 1x Digital Output |
40217-10613 | 6x Analog Outputs, 1x Digital Inputs, 3x Digital Outputs |
Need a Communication Protocol other than Modbus?
Be sure to mention this requirement when you request a quote and our Technical Sales Representatives will be sure to add the optional digital and analog I/O devices.